Photon and Neutron Dose Estimation Using Monte Carlo Simulation in TrueBeam’s Room

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • معلومة اضافية
    • بيانات النشر:
      Medknow, 2024.
    • الموضوع:
      2024
    • نبذة مختصرة :
      Purpose: The distribution of neutron ambient dose equivalent within the TrueBeam 10 MV photon chamber was investigated. Materials and Methods: The research used particle and heavy ion transport code system (PHITS) code and JENDL-5.0 to simulate the neutron ambient dose equivalent on and around TrueBeam’s head. The simulated results were compared with the measured results using CR-39 detectors when TrueBeam radiated 5000 monitor units of 10 MV photons with field sizes 20 cm × 20 cm and 0.5 cm × 0.5 cm. Results: Out of field size, the neutron ambient dose equivalents of the 0.5 cm × 0.5 cm field size are higher than those values of the 20 cm × 20 cm field size from 4% to 30%. The differences between the simulated value and the measured value of the neutron ambient dose equivalents at all points out of field size are smaller than 20%. Conclusion: The neutron ambient dose equivalents, simulated with PHITS and JENDL-5.0, are satisfied with the measured neutron ambient dose equivalent.
    • ISSN:
      1998-3913
      0971-6203
    • الرقم المعرف:
      10.4103/jmp.jmp_70_24
    • Rights:
      URL: http://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
    • الرقم المعرف:
      edsair.doi.dedup.....47854c1f99b98c5944f607b40bf4a266