Accuracy evaluation of CAD/CAM generated splints in orthognathic surgery: a cadaveric study.

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • معلومة اضافية
    • المصدر:
      Publisher: BioMed Central Country of Publication: England NLM ID: 101245792 Publication Model: Electronic Cited Medium: Internet ISSN: 1746-160X (Electronic) Linking ISSN: 1746160X NLM ISO Abbreviation: Head Face Med Subsets: MEDLINE
    • بيانات النشر:
      Original Publication: [London] : BioMed Central, [2005]-
    • الموضوع:
    • نبذة مختصرة :
      Introduction: To evaluate the accuracy of CAD/CAM generated splints in orthognathic surgery by comparing planned versus actual post-operative 3D images.
      Methods: Specific planning software (SimPlant(®) OMS Standalone 14.0) was used to perform a 3D virtual Le Fort I osteotomy in 10 fresh human cadaver heads. Stereolithographic splints were then generated and used during the surgical procedure to reposition the maxilla according to the planned position. Pre-operative planned and postoperative 3D CT scan images were fused and imported to dedicated software (MATLAB(®)) 7.11.) for calculating the translational and rotational (pitch, roll and yaw) differences between the two 3D images. Geometrical accuracy was estimated using the Root Mean Square Deviations (RMSD) and lower and upper limits of accuracy were computed using the Bland & Altman method, with 95 % confidence intervals around the limits. The accuracy cutoff was set at +/- 2 mm for translational and ≤ 4° for rotational measurements.
      Results: Overall accuracy between the two 3D images was within the accuracy cutoff for all values except for the antero-posterior positioning of the maxilla (2.17 mm). The translational and rotational differences due to the splint were all within the accuracy cutoff. However, the width of the limits of agreement (range between lower and upper limits) showed that rotational differences could be particularly large.
      Conclusion: This study demonstrated that maxillary repositioning can be accurately approximated and thus predicted by specific computational planning and CAD/CAM generated splints in orthognathic surgery. Further study should focus on the risk factors for inaccurate prediction.
    • References:
      J Oral Maxillofac Surg. 2013 Jan;71(1):128-42. (PMID: 22695016)
      Int J Oral Maxillofac Surg. 2013 May;42(5):592-6. (PMID: 23200425)
      J Oral Maxillofac Surg. 2013 May;71(5):911-20. (PMID: 23312847)
      J Oral Maxillofac Surg. 2013 Aug;71(8):1406-14. (PMID: 23642546)
      J Craniomaxillofac Surg. 2005 Feb;33(1):8-12. (PMID: 15694143)
      J Oral Maxillofac Surg. 2012 Oct;70(10):2419-26. (PMID: 22516840)
      Rev Stomatol Chir Maxillofac. 2004 Jun;105(3):165-9. (PMID: 15211215)
      J Oral Maxillofac Surg. 2003 Jul;61(7):814-7. (PMID: 12856256)
      J Oral Maxillofac Surg. 1997 Aug;55(8):811-6; discussion 817. (PMID: 9251608)
      J Craniomaxillofac Surg. 2006 Dec;34(8):478-83. (PMID: 17157519)
      Int J Oral Maxillofac Surg. 2007 Apr;36(4):296-300. (PMID: 17240117)
      Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008 Feb;105(2):e1-7. (PMID: 18230371)
      J Oral Maxillofac Surg. 2009 Aug;67(8):1628-39. (PMID: 19615574)
      J Oral Maxillofac Surg. 2009 Oct;67(10):2080-92. (PMID: 19761902)
      J Oral Maxillofac Surg. 2009 Oct;67(10):2093-106. (PMID: 19761903)
      Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010 Nov;110(5):e1-7. (PMID: 20813559)
      J Oral Maxillofac Surg. 2011 Jul;69(7):2014-24. (PMID: 21684451)
      J Craniomaxillofac Surg. 2012 Feb;40(2):162-8. (PMID: 21458285)
      Oral Surg Oral Med Oral Pathol Oral Radiol. 2012 May;113(5):673-87. (PMID: 22668627)
    • الموضوع:
      Date Created: 20150726 Date Completed: 20160108 Latest Revision: 20191210
    • الموضوع:
      20240829
    • الرقم المعرف:
      PMC4514936
    • الرقم المعرف:
      10.1186/s13005-015-0082-9
    • الرقم المعرف:
      26209339