نبذة مختصرة : OBJECTIVE: We evaluated the performance of the iLet bionic pancreas (BP) in non-Hispanic White individuals (here referred to as Whites) and in Black, Hispanic, and other individuals (here collectively referred to as Minorities). RESEARCH DESIGN AND METHODS: A multicenter, randomized controlled trial evaluated glycemic management with the BP versus standard of care (SC) in 161 adult and 165 pediatric participants with type 1 diabetes over 13 weeks. RESULTS: In Whites (n = 240), the mean baseline-adjusted difference in 13-week HbA1c between the BP and SC groups was -0.45% (95% CI -0.61 to -0.29 [-4.9 mmol/mol; -6.6 to -3.1]; P < 0.001), while this difference among Minorities (n = 84) was -0.53% (-0.83 to -0.24 [-6.0 mmol/mol; -9.2 to -2.8]; P < 0.001). In Whites, the mean baseline-adjusted difference in time in range between the BP and SC groups was 10% (95% CI 7-12; P < 0.001) and in Minorities was 14% (10-18; P < 0.001). CONCLUSIONS: The BP improves glycemic control in both Whites and Minorities and offers promise in decreasing health care disparities.