نبذة مختصرة : Computed tomographic coronary artery calcium scanning enables cardiovascular risk stratification; however, exposing patients to high radiation levels is an ongoing concern. New-generation computed tomographic systems use lower radiation doses than older systems do. To quantify comparative doses of radiation exposure, we prospectively acquired images from 220 patients with use of a 64-slice GE LightSpeed VCT scanner (control group, n=110) and a 256-slice GE Revolution scanner (study group, n=110). The groups were matched for age, sex, and body mass index; statistical analysis included t tests and linear regression. The mean dose-length product was 21% lower in the study group than in the control group (60.2 ± 27 vs 75.9 ± 22.6 mGy·cm; P
No Comments.