نبذة مختصرة : Background & aimWe studied FibroTouch (FT) and Fibroscan (FS) examination results and their repeatability when performed by healthcare personnel of different background.MethodsFT and FS examinations were performed on patients with chronic liver disease by two operators, a doctor and a nurse, twice on each patient, at two different time points, independent of each other.ResultsThe data for 163 patients with 1304 examinations was analyzed. There was strong correlation between FT and FS for attenuation parameter (Spearman's rho 0.76, pConclusionMeasurements obtained with FT and FS strongly correlated, but significant differences in their absolute values, consistency, examination time and number of invalid measurements were observed. Either device can be used by healthcare personnel of different backgrounds when sufficiently trained.