Ultra-High-Resolution and K-Edge Imaging of Prosthetic Heart Valves With Spectral Photon-Counting CT ; A Phantom Study

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • معلومة اضافية
    • بيانات النشر:
      Ovid Technologies (Wolters Kluwer Health)
    • الموضوع:
      2024
    • نبذة مختصرة :
      Background and Purpose The contribution of cardiac computed tomography (CT) for the detection and characterization of prosthetic heart valve (PHV) complications is still limited due mainly to artifacts. Computed tomography systems equipped with photon-counting detectors (PCDs) have the potential to overcome these limitations. Therefore, the aim of the study was to compare image quality of PHV with PCD-CT and dual-energy dual-layer CT (DEDL-CT). Materials and Methods Two metallic and 3 biological PHVs were placed in a tube containing diluted iodinated contrast inside a thoracic phantom and scanned repeatedly at different angles on a DEDL-CT and PCD-CT. Two small lesions (~2 mm thickness; containing muscle and fat, respectively) were attached to the structure of 4 valves, placed inside the thoracic phantom, with and without an extension ring, and scanned again. Acquisition parameters were matched for the 2 CT systems and used for all scans. Metallic valves were scanned again with parameters adapted for tungsten K-edge imaging. For all valves, different metallic parts were measured on conventional images to assess their thickness and blooming artifacts. In addition, 6 parallelepipeds per metallic valve were drawn, and all voxels with density <3 times the standard deviation of the contrast media were recorded as an estimate of streak artifacts. For subjective analysis, 3 expert readers assessed conventional images of the valves, with and without lesions, and tungsten K-edge images. Conspicuity and sharpness of the different parts of the valve, the lesions, metallic, and blooming artifacts were scored on a 4-point scale. Measurements and scores were compared with the paired t test or Wilcoxon test. Results The objective analysis showed that, with PCD-CT, valvular metallic structures were thinner and presented less blooming artifacts. Metallic artifacts were also reduced with PCD-CT (11 [interquartile (IQ) = 6] vs 40 [IQ = 13] % of voxels). Subjective analysis allowed noticing that some structures were visible or ...
    • الرقم المعرف:
      10.1097/rli.0000000000001068
    • الرقم المعرف:
      10.1097/RLI.0000000000001068
    • Rights:
      http://creativecommons.org/licenses/by-nc-nd/4.0/
    • الرقم المعرف:
      edsbas.FDA6EF44