نبذة مختصرة : Background: Malpositioning during total hip arthroplasty may cause dislocation, pain, and other complications. To evaluate the potential of sacral slope (SS) as a reliable parameter of pelvic flexion. Methods: We developed a model of pelvic flexion to determine the intraobserver and interobserver variability and reliability of SS measurements by lateral radiography by three independent observers. Results: Measurement error was 1.2 and the intraobserver reliability was moderate to substantial (Interclass correlation coefficient: 0.31 to 0.66). Based on the Spearman-Brown formula, the measurement is reliable if it is done at least seven times by two observers, and four times by three observers. Conclusions: The data suggest that measurement of SS of pelvic flexion is a clinically useful parameter for the optimization of THA conditions.