Agreement and repeatability of Icare ic100 tonometer

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • معلومة اضافية
    • بيانات النشر:
      Medknow, 2020.
    • الموضوع:
      2020
    • نبذة مختصرة :
      Purpose: To find the agreement and repeatability of Icare ic100 tonometer. Methods: We included 150 subjects above the age of 18 years for this cross-sectional, multicenter study with intraocular pressure (IOP) ≥7 mmHg. After the initial ophthalmic examination, two masked examiners took five IOP measurements using three different instruments; Icare ic100, Icare TA01i, and Goldmann applanation tonometer (GAT) in only one eye of the participants. Comparison of agreement of IOP using different instruments was quantified with intraclass correlation coefficient (ICC) using the two-way random effects models of absolute agreement and Cronbach's alpha. The test-retest variability of the instruments was assessed by deriving repeatability coefficient (RC) and coefficient of variation (CV). Results: Agreement between the tonometers across the different IOP groups had no statistically significant difference in their mean IOP. Icare ic100 was found to have good reliability across all IOP groups (ICC value >0.78) when compared with Icare TA01i. In comparison with GAT, Icare ic100 showed good reliability across all IOP groups (ICC >0.87) except >16 to