Influence of maximum MLC leaf speed on the quality of volumetric modulated arc therapy plans

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • معلومة اضافية
    • بيانات النشر:
      Wiley, 2020.
    • الموضوع:
      2020
    • نبذة مختصرة :
      Purpose Maximum leaf speed is a configurable parameter of MLC in a treatment‐planning system. This study investigated the influence of MLC on the quality of VMAT plans. Methods Seven MLCs with different maximum leaf speeds (1.0, 1.5, 2.25, 3.5, 5.0, 7.5, and 10 cm/s) were configured for an accelerator in treatment‐planning system. Correspondingly, seven treatment plans, with the identical initial optimization parameter, were designed with the mdaccAutoPlan system. Six nasopharyngeal carcinoma (NPC) patients and nine rectal cancer patients were selected, representing complex and simple clinical circumstances. VMAT plan quality was evaluated with PlanIQTM software. The results were statistically analyzed with a one‐way analysis of variance (ANOVA) and pairwise comparison tests. Results The relative changes of plan scores achieved by the seven configured accelerators, with specific maximum MLC leaf speed (MMLS) for each patient, were studied. Two apparent trends of MMLS influence on VMAT plan scores were observed: Plan scores increased with MMLS; Plan scores increased rapidly when MMLS increased from 1 to 3.5, thus the relative change of plan score decreased in this MMLS range. The stationary point of maximum MLC speed (MMSSP) is defined, for the specific MMLS when the relative changes of plan scores is first
    • ISSN:
      1526-9914
    • الرقم المعرف:
      10.1002/acm2.13020
    • Rights:
      OPEN
    • الرقم المعرف:
      edsair.doi.dedup.....cced1821106f924a79edc53947db0ad0