Local impedance‐guided radiofrequency ablation with standard and high power: Results of a preclinical investigation

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • معلومة اضافية
    • بيانات النشر:
      Wiley, 2021.
    • الموضوع:
      2021
    • نبذة مختصرة :
      BACKGROUND Local impedance (LI) drop measured with microfidelity electrodes embedded in the tip of an ablation catheter accurately reflects tissue heating during radiofrequency (RF) ablation. Previous studies found 15-30 Ω LI drops created successful lesions, while more than 40 Ω drops were associated with steam pops. The objective of this study was to evaluate the safety and efficacy of LI-guided ablation using standard (30 W) and high-power (50 W) in a preclinical model. METHODS RF lesions were created in explanted swine hearts (n = 6) to assess the feasibility of LI-guided ablation by targeting 10, 20, or 30 Ω (n = 20/group) drops. Subsequently, LI-guided ablation was evaluated in a chronic animal model (n = 8 Canines, 25-29 kg, 30/50 W). During the index procedure point-by-point intercaval line ablation and left inferior pulmonary vein (PV) isolation were performed. RF duration was at the operators' discretion but discontinued early if a 15-30 Ω drop was achieved. Operators attempted to avoid LI drops of more than 40 Ω. At 1-month, durable conduction block was evaluated with electroanatomic mapping followed by necropsy and histopathology. RESULTS In explanted tissue, terminating ablation at 10, 20, or 30 Ω LI drops created statistically larger lesions (p
    • ISSN:
      1540-8167
      1045-3873
    • Rights:
      CLOSED
    • الرقم المعرف:
      edsair.doi.dedup.....c7abbb77188f3fcc52f5d02e69b8b5f9