Finite element analysis of new headless compression supporting screw for the treatment of unstable femoral neck fracture

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • معلومة اضافية
    • بيانات النشر:
      IOS Press, 2022.
    • الموضوع:
      2022
    • نبذة مختصرة :
      BACKGROUND: Femoral neck fracture is an unsolved challenge in orthopedics. The complication rate in particular is high. There remains a lack of consensus on the optimal choice of internal fixation for unstable femoral neck fracture. OBJECTIVE: The study aimed to develop a new headless compression supporting screw (HCSS) for the treatment of unstable foemoral neck fracture. METHODS: We designed a new HCSS and used a femoral neck fracture (Pauwels III fracture) model (left, fourth-generation composite, Sawbones) and three-dimensional finite element analysis to compare the biomechanical performance of HCSSs with that of cannulated compression screws (CCSs) for treatment of unstable femoral neck fracture. RESULTS: Maximum displacement, peak von Mises stress, peak strain, and rotation for the HCSS were smaller than those for the CCS. The stress was more widely distributed for the HCSS, whereas the stress was concentrated for the CCS. CONCLUSIONS: The HCSS resulted in better biomechanical stability than that from the CCS. For Pauwels III fractures the HCSS exhibits better resistance to shear forces and better support, providing a new clinical treatment.