Item request has been placed! ×
Item request cannot be made. ×
loading  Processing Request

Reverse hybrid and cemented hip replacement compared using radiostereometry and dual-energy X-ray absorptiometry

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • معلومة اضافية
    • بيانات النشر:
      Medical Journals Sweden AB, 2012.
    • الموضوع:
      2012
    • نبذة مختصرة :
      Background and purpose Total hip replacement (THR) with a reverse hybrid (RH), a combination of a cemented polyethylene cup and a cementless femoral stem, has been increasingly used in Scandinavia. In a randomized trial, we compared an RH THR with a proximal hydroxyapatite- (HA-) coated stem to a conventional cemented THR. Both groups received the same polyethylene cup. Patients and methods 51 patients (52 hips) were included. Radiostereometry (RSA) and dual-energy X-ray absorptiometry (DEXA) were performed postoperatively and after 6, 12, and 24 months. 42 patients (43 hips) were followed for 2 years. Results Mean cup rotation around the x-axis was 0.13° for the cemented group and –0.24° for the RH group (p = 0.03). Cup migration in the other axes, and stem migration and wear were similar between the 2 study groups. Bone remodeling around the cup was also similar between the groups. Bone loss in Gruen zone 1 was 18% for the cementless stems, as compared to an increase of 1.4% for the cemented ones (p < 0.001). Bone loss was similar in the other Gruen zones. Harris hip score and Oxford hip score were similar pre- and postoperatively in the 2 groups. Interpretation In the present study, RH THR with a cementless hydroxyapatite-coated stem and conventional cemented THR did not show any major differences regarding stem migration and bone loss after 2 years of follow-up.
    • ISSN:
      1745-3682
      1745-3674
    • Rights:
      OPEN
    • الرقم المعرف:
      edsair.doi.dedup.....5d1202c1a28dd5300684ac1bb89b58f7