Item request has been placed! ×
Item request cannot be made. ×
loading  Processing Request

Effects of Exposure Parameters and Voxel Size for Cone-Beam Computed Tomography on the Image Matching Accuracy with an Optical Dental Scan Image: An In Vitro Study

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • معلومة اضافية
    • بيانات النشر:
      Hindawi Limited, 2021.
    • الموضوع:
      2021
    • نبذة مختصرة :
      This study is aimed at assessing the effects of exposure parameters and voxel size for cone-beam computed tomography (CBCT) on the image matching accuracy with an optical dental scan image. CBCT and optical scan images of a dry human mandible were obtained. Different CBCT settings were used: tube voltage, 60, 80, and 100 kVp; tube current, 6 and 8 mA; and voxel size, 100, 200, and 300 μm. Image matching between the CBCT and optical scan images was performed using implant planning software by dental professionals ( n = 18 ). The image matching accuracy in each combination of CBCT settings was evaluated by assessing the linear discrepancy between the three-dimensionally reconstructed radiological image and the registered optical scan image using an image analysis software program. The Kruskal-Wallis test and a post hoc Mann–Whitney U test with Bonferroni correction were used to compare the accuracy of image registration between the groups ( α = 0.05 ). Overall, the image matching accuracy was not significantly different between tube voltage and current settings; however, significantly higher image registration errors were found at the combination of 100 kVp tube voltage/8 mA tube current ( F = 8.44 , P < 0.001 ). Changes in voxel sizes did not significantly interfere with the image registration results. No interaction was found among voltage, current, and voxel size in terms of image registration accuracy ( F = 2.022 , P = 0.091 ). Different exposure parameter settings in tube voltage and tube current did not significantly influence the image matching accuracy between CBCT and optical dental scan images; however, a high radiation dose could be inappropriate. The image matching accuracy was not significantly affected by changing the voxel sizes of CBCT.
    • File Description:
      text/xhtml
    • ISSN:
      2314-6141
      2314-6133
    • Rights:
      OPEN
    • الرقم المعرف:
      edsair.doi.dedup.....534ec5f58b856d40a87762759bbed87f