Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • معلومة اضافية
    • بيانات النشر:
      Hindawi Limited, 2016.
    • الموضوع:
      2016
    • نبذة مختصرة :
      Background.In premature infants, maintaining blood partial pressure of carbon dioxide (pCO2) value within a narrow range is important to avoid cerebral lesions. The aim of this study was to assess the accuracy of a noninvasive transcutaneous method (TcpCO2), compared to blood partial pressure of carbon dioxide (pCO2).Methods.Retrospective observational study in a tertiary neonatal intensive care unit. We analyzed the correlation between blood pCO2and transcutaneous values and the accuracy between the trends of blood pCO2and TcpCO2in all consecutive premature infants born at Results.248 infants were included (median gestational age: 29 + 5 weeks and median birth weight: 1250 g), providing 1365 pairs of TcpCO2and blood pCO2values. Pearson’sRcorrelation between these values was 0.58. The mean bias was −0.93 kPa with a 95% confidence limit of agreement of −4.05 to +2.16 kPa. Correlation between the trends of TcpCO2and blood pCO2values was good in only 39.6%.Conclusions.In premature infants, TcpCO2was poorly correlated to blood pCO2, with a wide limit of agreement. Furthermore, concordance between trends was equally low. We warn about clinical decision-making on TcpCO2alone when used as continuous monitoring.
    • File Description:
      text/xhtml
    • ISSN:
      2090-1313
      2090-1305
    • Rights:
      OPEN
    • الرقم المعرف:
      edsair.doi.dedup.....17558a808720edd7b509c1318e837ec7