نبذة مختصرة : The patients who had been operated total mastectomy are different from general women in their breasts thickness due to breast surgery. As a result, digital chest image from total mastectomy patients will be different attenuation. The main objective for this study is to show that a proper Ion chamber standard combination measuring MTF which is objective basis for Digital image, when be x-ray for total mastectomy patient. We have designed the unique number that shown Left is 1, Right is 2, Center is 3 and have put the edge phantom on detector ion chamber. Lastly, we have obtained experiment images. The evaluations of all image quality have measured by 50% MTF of spatial resolution and absorption dose using Matlab(R2007a). The result showed that average exposure condition, MTF value, absorption dose for 1+3 and 2+3 combinations were 2.745 mAs, 1.925 lp/mm, 0.688 mGy. Consequently, that showed high MTF, DQE and low dose than other combinations. Therefore, a proper changes of ion chambers are able to improve image quality and to reduce radiation exposure when be X-ray for total mastectomy patients. Also, it will be possible to standard for application chamber combination and utilization on clinical detection.