Respirometry: Correcting for Diffusion and Validating the Use of Plastic Multiwell Plates with Integrated Optodes.

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • معلومة اضافية
    • نبذة مختصرة :
      Polystyrene multiwell plates with integrated optodes act as multiple closed-system respirometers that enable the simultaneous measurement of oxygen consumption in small animals. However, the diffusion of oxygen through polystyrene needs to be taken into consideration. Here we provide an equation that accounts for the empirically determined rate of oxygen through a polystyrene well when calculating the instantaneous rate of oxygen consumption. Furthermore, we describe a novel method of calibrating a small respirometer for accuracy using micro-osmotic pumps containing an oxygen scavenger, which is delivered at a constant rate and therefore yields a constant rate of oxygen consumption in an airtight system. [ABSTRACT FROM AUTHOR]
    • نبذة مختصرة :
      Copyright of Physiological & Biochemical Zoology is the property of University of Chicago Press and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)