Item request has been placed! ×
Item request cannot be made. ×
loading  Processing Request

Design of a lightweight passive orthosis for tremor suppression.

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • معلومة اضافية
    • المصدر:
      Publisher: BioMed Central Country of Publication: England NLM ID: 101232233 Publication Model: Electronic Cited Medium: Internet ISSN: 1743-0003 (Electronic) Linking ISSN: 17430003 NLM ISO Abbreviation: J Neuroeng Rehabil Subsets: MEDLINE
    • بيانات النشر:
      Original Publication: [London] : BioMed Central, 2004-
    • الموضوع:
    • نبذة مختصرة :
      Background: Tremor is the most common movement disorder with the highest prevalence in the upper limbs. The mechanical suppression of involuntary movements is an alternative and additional treatment to medication or surgery. Here we present a new, soft, lightweight, task asjustable and passive orthosis for tremor suppression.
      Methods: A new concept of a manual, textile-based, passive orthosis was designed with an integrated, task adjustable, air-filled structure, which can easily be inflated or deflated on-demand for a certain daily activity. The air-filled structure is placed on the dorsal side of the wrist and gets bent and compressed by movements when inflated. In a constant volume air-filled structure, air pressure increases while it is inflating, creating a counterforce to the compression caused by bending. We characterised the air-filled structure stiffness by measuring the reaction torque as a function of the angle of deflection on a test bench. Furthermore, we evaluated the efficacy of the developed passive soft orthosis by analysing the suppression of involuntary movements in the wrist of a tremor-affected patient during different activities of daily living (i.e. by calculating the power spectral densities of acceleration).
      Results: By putting special emphasis on the comfort and wearability of the orthosis, we achieved a lightweight design (33 g). The measurements of the angular deflection and resulting reaction torques show non-linear, hysteretic, behaviour, as well as linear behaviour with a coefficient of determination (R 2 ) between 0.95 and 0.99. Furthermore, we demonstrated that the soft orthosis significantly reduces tremor power for daily living activities, such as drinking from a cup, pouring water and drawing a spiral, by 74 to 82% (p = 0.03); confirmed by subjective tremor-reducing perception by the patient.
      Conclusion: The orthosis we developed is a lightweight and unobtrusive assistive technology, which suppresses involuntary movements and shows high wearability properties, with the potential to be comfortable. This air-structure technology could also be applied to other movement disorders, like spasticity, or even be integrated into future exoskeletons and exosuits for the implementation of variable stiffness in the systems.
    • References:
      Ann Rheum Dis. 1968 Mar;27(2):175-7. (PMID: 5646436)
      Mov Disord. 1990;5(2):118-26. (PMID: 2325673)
      Sci Technol Adv Mater. 2015 Sep 25;16(5):055002. (PMID: 27877837)
      Med Biol Eng Comput. 2018 May;56(5):923-930. (PMID: 29101536)
      Am J Occup Ther. 2016 Jan-Feb;70(1):7001350010p1-7001350010p10. (PMID: 26709433)
      J Neuroeng Rehabil. 2018 Feb 17;15(1):9. (PMID: 29454392)
      Semin Musculoskelet Radiol. 2001 Sep;5(3):235-40. (PMID: 11595967)
      Mov Disord. 2007 Apr 15;22(5):666-72. (PMID: 17266084)
      J Rehabil Res Dev. 2016;53(6):1151-1166. (PMID: 28355037)
      Parkinsonism Relat Disord. 2011 Aug;17(7):509-15. (PMID: 21632272)
      J Neurol. 1993 Nov;241(1):54-61. (PMID: 8138823)
      J Electromyogr Kinesiol. 2014 Apr;24(2):300-6. (PMID: 24462391)
      Ann Neurol. 2018 Jan;83(1):107-114. (PMID: 29265546)
      J Bodyw Mov Ther. 2013 Oct;17(4):488-94. (PMID: 24139008)
      Tremor Other Hyperkinet Mov (N Y). 2012;2:. (PMID: 23439994)
      Parkinsonism Relat Disord. 2010 Nov;16(9):604-7. (PMID: 20691629)
      Stereotact Funct Neurosurg. 2018;96(1):54-59. (PMID: 29408817)
      Stud Health Technol Inform. 2017;242:741-747. (PMID: 28873879)
      Mov Disord. 2010 Apr 15;25(5):534-41. (PMID: 20175185)
      Muscle Nerve. 2001 Jun;24(6):716-35. (PMID: 11360255)
      J Neurosurg. 2018 Apr;128(4):1214-1221. (PMID: 28665251)
      Parkinsonism Relat Disord. 2011 Aug;17(7):537-42. (PMID: 21570891)
      Mov Disord. 2011 May;26(6):1096-105. (PMID: 21626554)
      Mov Disord. 1998;13 Suppl 3:2-23. (PMID: 9827589)
      Mov Disord. 2018 Jan;33(1):75-87. (PMID: 29193359)
      Mov Disord. 2000 Jan;15(1):84-94. (PMID: 10634246)
      J Neurophysiol. 2012 Aug;108(4):1158-66. (PMID: 22649208)
      Ind Health. 2010;48(1):108-14. (PMID: 20160414)
      Mov Disord. 2013 Nov;28(13):1793-800. (PMID: 24038576)
      Ann Rheum Dis. 2003 Feb;62(2):151-8. (PMID: 12525385)
      Ergonomics. 1988 Jan;31(1):97-131. (PMID: 3359991)
      Neurology. 1995 Dec;45(12):2143-6. (PMID: 8848182)
      Arch Phys Med Rehabil. 2005 Aug;86(8):1661-4. (PMID: 16084823)
      J Neuroeng Rehabil. 2019 Jul 18;16(1):93. (PMID: 31319893)
      Tremor Other Hyperkinet Mov (N Y). 2014 Aug 14;4:259. (PMID: 25157323)
    • Contributed Indexing:
      Keywords: Comfort; Laser welding; Orthosis; Soft; Suppression; Textile integrated; Tremor; Upper limb; Variable stiffness; Wearable
    • الموضوع:
      Date Created: 20200411 Date Completed: 20201027 Latest Revision: 20201027
    • الموضوع:
      20231215
    • الرقم المعرف:
      PMC7144059
    • الرقم المعرف:
      10.1186/s12984-020-00673-7
    • الرقم المعرف:
      32272964