Dental Implants Fatigue as a Possible Failure of Implantologic Treatment: The Importance of Randomness in Fatigue Behaviour.

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • معلومة اضافية
    • المصدر:
      Publisher: Wiley Country of Publication: United States NLM ID: 101600173 Publication Model: Print-Electronic Cited Medium: Internet ISSN: 2314-6141 (Electronic) NLM ISO Abbreviation: Biomed Res Int Subsets: MEDLINE
    • بيانات النشر:
      Publication: 2024- : [Hoboken, NJ] : Wiley
      Original Publication: New York, NY : Hindawi Pub. Co.
    • الموضوع:
    • نبذة مختصرة :
      Objective: To show how random variables concern fatigue behaviour by a probabilistic finite element method.
      Methods: Uncertainties on material properties due to the existence of defects that cause material elastic constant are not the same in the whole dental implant the dimensions of the structural element and load history have a decisive influence on the fatigue process and therefore on the life of a dental implant. In order to measure these uncertainties, we used a method based on Markoff chains, Bogdanoff and Kozin cumulative damage model, and probabilistic finite elements method.
      Results: The results have been obtained by conventional and probabilistic methods. Mathematical models obtained the same result regarding fatigue life; however, the probabilistic model obtained a greater mean life but with more information because of the cumulative probability function.
      Conclusions: The present paper introduces an improved procedure to study fatigue behaviour in order to know statistics of the fatigue life (mean and variance) and its probability of failure (fatigue life versus probability of failure).
    • References:
      J Oral Rehabil. 2010 Jul;37(7):497-500. (PMID: 20236234)
      J Oral Rehabil. 2011 Apr;38(4):286-94. (PMID: 20969613)
      Comput Methods Biomech Biomed Engin. 2012;15(3):263-73. (PMID: 21170768)
      J Oral Rehabil. 2012 Jul;39(7):489-501. (PMID: 22489928)
      Biomaterials. 2010 Mar;31(7):1465-85. (PMID: 19969343)
      J Oral Rehabil. 2008 Jul;35(7):495-508. (PMID: 18557916)
      Acta Orthop Scand. 1981;52(2):155-70. (PMID: 7246093)
      Arch Oral Biol. 1985;30(7):563-9. (PMID: 3864406)
      Clin Oral Implants Res. 2005 Aug;16(4):486-94. (PMID: 16117775)
      J Oral Rehabil. 2006 Feb;33(2):152-9. (PMID: 16457676)
      J Oral Rehabil. 2006 Apr;33(4):293-300. (PMID: 16629884)
      Clin Oral Implants Res. 1998 Dec;9(6):407-18. (PMID: 11429942)
    • الرقم المعرف:
      0 (Dental Implants)
      0 (Dental Materials)
      D1JT611TNE (Titanium)
    • الموضوع:
      Date Created: 20151120 Date Completed: 20160830 Latest Revision: 20181113
    • الموضوع:
      20250114
    • الرقم المعرف:
      PMC4637060
    • الرقم المعرف:
      10.1155/2015/825402
    • الرقم المعرف:
      26583137